8. Transferring measurements
Connect the blood pressure monitor to your PC using the
USB cable.
No data transfer may be launched whilst performing a
measurement.
PC is shown on the display. Begin the data transfer
in the “HealthManager” PC software. During the
data transfer, an animation is shown on the
display. A successful data transfer is displayed
as in figure 1. If the data transfer is unsuccessful,
an error message appears as in figure 2. In this
case, interrupt the PC connection and start the
data transfer again.
After 30 seconds of not being in use or if com-
munication with the PC is interrupted, the blood
pressure monitor switches itself off automati-
cally
.
9. Error messages/trouble-shooting
In case of faults, the _ message appears in the display.
Error messages may appear if:
systolic or diastolic pressure could not be measured (
or appears on the display)
systolic or diastolic pressure was outside the measure-
ment range (
or Lo appears on the display)
the cuff is fastened too tightly or loosely (
or
appears on the display)
the pump pressure is higher than 300 mmHg (
appears on the display)
pumping up takes longer than 160 seconds ( appears
on the display)
there is a system or device error ( , , or
appears on the display)
the batteries are almost empty .
In such cases, repeat the measurement. Ensure you do not
move or speak.
If necessary, re-insert or replace the batteries.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section “Technical speci-
fications”, the technical alarm will appear on the display
indicating either “
” or “Lo. In such cases, you should seek
medical assistance and check the accuracy of your proce-
dure.
The limit values for the technical alarm are factory set and
cannot be adjusted or deactivated. These alarm limit val-
ues are accorded second priority under the standard IEC
60601-1-8.
A
bb. 1
A
bb. 2